pH Range |
|
basic mode: -2.00 to 16.00 pH; standard mode: -2.000 to 16.000 pH |
Resolution |
|
basic mode: 0.01 pH; standard mode: 0.001 pH |
Accuracy (@25°C) |
|
basic mode: ±0.01 pH; standard mode: ±0.002 pH |
Calibration |
|
basic mode: automatic, up to three points calibration, 5 standard buffers available; standard mode: automatic, up to five point calibration, 7 standard buffers available and two custom buffers |
Temperature Compensation |
|
automatic, -5.0 to 100.0°C (using integral temperature sensor) |
Electrode Diagnostics |
|
standard mode: probe condition, response time and out of calibration range |
mV pH Range |
|
±1000 mV |
Resolution |
|
0.1 mV |
Accuracy (@25°C) |
|
±0.2 mV |
ORP Range |
|
±2000 mV |
Resolution |
|
0.1 mV |
Accuracy (@25°C) |
|
±0.2 mV |
Temperature Range |
|
-20.0 to 120.0°C |
Resolution |
|
0.1°C |
Accuracy |
|
±0.5°C |
Additional Specifications: |
|
|
Probe |
|
HI-11310 digital glass body pH electrode with 3.5mm connector and 1m cable |
Logging |
|
up to 1000 (400 for basic mode) records organised in: manual log-on demand (max. 200 logs), manual log-on-stability (max. 200 logs), interval logging (max. 600 samples; 100 lots) |
Connectivity |
|
1 USB port for storage; 1 micro USB port for charging and PC connectivity |
Environment |
|
0 to 50°C; RH max 95% non-condensing |
Power Supply |
|
5 VDC adapter (included) |
Dimensions |
|
2020 x 140 x 12 mm |
Weight |
|
250 g |